SDS1000CFL Service Manual 27
the Vpp measurement of the oscilloscope is approximately 750mv.
16. Press Trigger LEVEL button to move the waveform to the screen
center.
17. Adjust the Trigger LEVEL within the waveform range to check if
there is stable trigger.
18. Disconnect the test connection.
19. Check the other channels in the same way as step 1 to 18.