38 SDS1000CFL Service Manual
Removing the Metal Shelf and Interface Module
Figure 5-7 Removing the Metal Shelf and Interface module
Removing steps:
1. Place the oscilloscope face down on a soft surface such as an anti-static
mat.
2. Disconnect the interface cable from the main board module.
3. Remove the four screws located on the interface module.
4. Remove the six screws located on the edge of the metal shelf.
5. Separate the interface module from the bottom of the oscilloscope and the
metal shelf from the front panel, then lift it up and off carefully to avoid
personal injury from the sharp edge of the metal shelf.
To install the rear panel, please operate as the reverse steps.
Interface module screw
Metal shelf screw
Metal shelf