30
SDS2000X HD Quick Start
Quick Action
Push the
button on the front panel to perform a specified action. Touch
Utility
>
Quick Action to open the Quick Action menu and set the action
mapping to the button. By default, Quick Action is disabled, and pushing the button
for the first time will open the Quick Action menu.
Calibration
Self Calibration
The oscilloscope is calibrated at the factory prior to being shipped. The calibration
is run at 23 °C (± 2 °C) and is valid for temperatures 23 ± 5 °C. Within this
temperature range, the oscilloscope will meet all specifications once warmed up.
Touch Utility
>
Menu
>
Maintenance
>
Self Calibration
Warm up the oscilloscope for at least 20 minutes prior to each use or calibration in
order for it to reach a stable operating temperature. Specifications are not
guaranteed during the warm-up period.
Whenever the oscilloscope is used in an environment outside 23 ± 5 °C, or when it
has been more than one month since the previous calibration, a self-calibration is
recommended.
Quick Cal
When Quick Cal is enabled, the oscilloscope will initiate an ADC calibration
procedure when the ADC temperature changes by over 3 °C to avoid performance
drift. This procedure is quick enough not to affect the user experience but will
interrupt the acquisition procedure. It is recommended to disable it when capturing
long-duration waveforms or monitoring rare events without interruption. The default
status of Quick Cal is disabled.