Note:
Please refer to the cursors chapter for more information.
Note:
Signals with DC components or deviation would cause error or deviation of the
FFT waveform components. To reduce the DC components, set the Channel
Coupling
to
AC.
To reduce the random noise and aliasing frequency components of repetitive or
single pulses, set the
Acquisition
of the oscilloscope to
Average
.
Figure 56 FFT Waveform in Split Mode
11. Press the
AUTO SET
softkey to automatically set the appropriate parameters for the
FFT measurement.
12. Press the
Display
softkey and select
Normal
,
Maximum
Hold
and
Average
to
display the FFT waveform.
To measure FFT waveform:
To make cursor measurements, press the
Cursors
button, and then press the
Mode
softkey to select
On
to turn the cursors. Use the X1 and X2 cursors to measure frequency
values and the difference between two frequency values (ΔX). Use the Y1 and Y2 cursors
to measure amplitude in dB and difference in amplitude (ΔY).
You can find the frequency value at the first occurrence of the waveform maximum by
using the X at Max Y measurement.
Summary of Contents for SDS2000X-E Series
Page 1: ...User Manual SDS2000X E Series Digital Oscilloscope UM0102E E01A SIGLENT TECHNOLOGIES CO LTD...
Page 14: ......
Page 69: ...SIGLENT SDS2000X E User Manual 43 Figure 13 x Interpolation Figure 14 Sin x x Interpolation...
Page 98: ...SIGLENT 72 SDS2000X E User Manual Figure 29 Relative Window Trigger...
Page 100: ...SIGLENT 74 SDS2000X E User Manual Figure 30 Interval Trigger...
Page 103: ...SIGLENT SDS2000X E User Manual 77 Figure 32 State Dropout Trigger...
Page 105: ...SIGLENT SDS2000X E User Manual 79 Figure 33 Runt Trigger...
Page 125: ...SIGLENT SDS2000X E User Manual 99 Figure 49 CAN Trigger...
Page 129: ...SIGLENT SDS2000X E User Manual 103 Figure 51 LIN Trigger...
Page 162: ...SIGLENT 136 SDS2000X E User Manual...
Page 200: ...SIGLENT 174 SDS2000X E User Manual Figure 95 Built in Arb interface...
Page 206: ...SIGLENT 180 SDS2000X E User Manual Figure 100 WiFi setting menu...
Page 215: ...SIGLENT SDS2000X E User Manual 189 Figure 108 Option Information...