SIG
L
ENT
122
SDS2000X-E
User Manual
Type of Measurements
Voltage Measurements
Voltage measurements include 17 kinds of voltage parameter measurements.
Figure 62 Voltage Measurements
1. Peak-Peak:
Difference between maximum and minimum data values.
2. Maximum:
Highest value of the input waveform.
3. Minimum:
Lowest value of the input waveform.
4. Amplitude:
Difference between the top and the base in a bimodal signal, or between
max and min in a unimodal signal.
5. Top:
Value of most probable higher state in a bimodal waveform.
6. Base:
Value of most probable lower state in a bimodal waveform.
7. Mean:
Average of data values in the first cycle.
8. Cycle mean:
Average of all data values.
9. Stdev:
Standard deviation of all data values.
10. Cycle Stdev:
Standard deviation of all data values in the first cycle.
11. RMS:
Root mean square of all data values.
12. Cycle RMS:
Root mean square of all data values in the first cycle.
13. Overshoot:
Overshoot is distortion that follows a major edge transition expressed as
a percentage of amplitude. ROV means rising edge overshoot and FOV means falling
edge overshoot.
Summary of Contents for SDS2000X-E Series
Page 1: ...User Manual SDS2000X E Series Digital Oscilloscope UM0102E E01A SIGLENT TECHNOLOGIES CO LTD...
Page 14: ......
Page 69: ...SIGLENT SDS2000X E User Manual 43 Figure 13 x Interpolation Figure 14 Sin x x Interpolation...
Page 98: ...SIGLENT 72 SDS2000X E User Manual Figure 29 Relative Window Trigger...
Page 100: ...SIGLENT 74 SDS2000X E User Manual Figure 30 Interval Trigger...
Page 103: ...SIGLENT SDS2000X E User Manual 77 Figure 32 State Dropout Trigger...
Page 105: ...SIGLENT SDS2000X E User Manual 79 Figure 33 Runt Trigger...
Page 125: ...SIGLENT SDS2000X E User Manual 99 Figure 49 CAN Trigger...
Page 129: ...SIGLENT SDS2000X E User Manual 103 Figure 51 LIN Trigger...
Page 162: ...SIGLENT 136 SDS2000X E User Manual...
Page 200: ...SIGLENT 174 SDS2000X E User Manual Figure 95 Built in Arb interface...
Page 206: ...SIGLENT 180 SDS2000X E User Manual Figure 100 WiFi setting menu...
Page 215: ...SIGLENT SDS2000X E User Manual 189 Figure 108 Option Information...