SIG
L
ENT
54
SDS2000X-E
User Manual
Trigger Source
The oscilloscope’s trigger source includes analog channels,
EXT
,
EXT/5
and the
AC Line
.
Press the
Setup
button on the front panel to enter the TRIGGER function menu; press the
Source
softkey and then turn the
Universal Knob
to select the desired trigger source.
The current trigger source is displayed at the upper right corner of the screen. Select
channel with signal input as trigger source to obtain stable trigger.
Analog channel input:
Signals input from analog channels can all be used as the trigger source.
External trigger input:
The external trigger source can be used to connect an external trigger signal to the EXT
TRIG channel, when the oscilloscope collects data simultaneously on multiple analog
channels. The trigger signal (such as an external clock/signal of the circuit to be tested)
should be connected to
EXT
and
EXT/5
trigger source via the [EXT TRIG] connector.
EXT/5 trigger source attenuates the signal by a factor of 5. It extends the trigger level. You
can set the trigger condition within the range of the trigger level (-600 mV to +600 mV).
AC line:
The trigger signal is obtained from the AC power input of the oscilloscope. This kind of
trigger can be used to display the relationship between a signal (such as an illuminating
device) and power (power supply device). For example, it is mainly used in the power
industry to stably trigger the waveform output from the transformer of a transformer
substation.
Note:
Select a stable channel waveform as the trigger source to stabilize the display.
Summary of Contents for SDS2000X-E Series
Page 1: ...User Manual SDS2000X E Series Digital Oscilloscope UM0102E E01A SIGLENT TECHNOLOGIES CO LTD...
Page 14: ......
Page 69: ...SIGLENT SDS2000X E User Manual 43 Figure 13 x Interpolation Figure 14 Sin x x Interpolation...
Page 98: ...SIGLENT 72 SDS2000X E User Manual Figure 29 Relative Window Trigger...
Page 100: ...SIGLENT 74 SDS2000X E User Manual Figure 30 Interval Trigger...
Page 103: ...SIGLENT SDS2000X E User Manual 77 Figure 32 State Dropout Trigger...
Page 105: ...SIGLENT SDS2000X E User Manual 79 Figure 33 Runt Trigger...
Page 125: ...SIGLENT SDS2000X E User Manual 99 Figure 49 CAN Trigger...
Page 129: ...SIGLENT SDS2000X E User Manual 103 Figure 51 LIN Trigger...
Page 162: ...SIGLENT 136 SDS2000X E User Manual...
Page 200: ...SIGLENT 174 SDS2000X E User Manual Figure 95 Built in Arb interface...
Page 206: ...SIGLENT 180 SDS2000X E User Manual Figure 100 WiFi setting menu...
Page 215: ...SIGLENT SDS2000X E User Manual 189 Figure 108 Option Information...