SIG
L
ENT
SDS2000X-E User Manual
101
LIN Trigger and Serial Decode
The following section describes “
Setup for LIN Signals
”, “
LIN Trigger
” and “
LIN Serial
Decode
” and includes instructions ontriggering and decoding LIN signals.
Setup for LIN Signals
There are two steps of setting the LIN signal: connecting the signal to oscilloscope and
specifying the parameters of each input signal.
1. Press the
Decode
key to enter the DECODE function menu.
2. Press the
Decode
softkey and select the desired slot (Serial 1 or Serial 2).
3. Press
Protocol
softkey and then select
LIN
by turning
Universal Knob
.
4. Press
Signal
softkey to enter the
SIGNAL
menu as Figure 50 shows.
Figure 50 LIN SIGNAL Menu
5. Press
Source
softkey to select the channel that is connected to the LIN signal.
6. Press the
Threshold
softkey
and set the LIN signal’s threshold voltage level by the
Universal Knob
. The threshold voltage level is for decoding, and it will be used as
the trigger voltage level when the trigger type is set to serial.
7. Press
softkey to return previous menu.
8. Press the
Configure
softkey to enter the
BUS CONFIG
menu.
9. Press
Baud
softkey to set baud rate.
The baud rate can be set as predefined value.
If the desired baud rate is not listed, select
custom
option, press the
Custom
and turn the
Universal Knob
to set the desired baud rate.
Summary of Contents for SDS2000X-E Series
Page 1: ...User Manual SDS2000X E Series Digital Oscilloscope UM0102E E01A SIGLENT TECHNOLOGIES CO LTD...
Page 14: ......
Page 69: ...SIGLENT SDS2000X E User Manual 43 Figure 13 x Interpolation Figure 14 Sin x x Interpolation...
Page 98: ...SIGLENT 72 SDS2000X E User Manual Figure 29 Relative Window Trigger...
Page 100: ...SIGLENT 74 SDS2000X E User Manual Figure 30 Interval Trigger...
Page 103: ...SIGLENT SDS2000X E User Manual 77 Figure 32 State Dropout Trigger...
Page 105: ...SIGLENT SDS2000X E User Manual 79 Figure 33 Runt Trigger...
Page 125: ...SIGLENT SDS2000X E User Manual 99 Figure 49 CAN Trigger...
Page 129: ...SIGLENT SDS2000X E User Manual 103 Figure 51 LIN Trigger...
Page 162: ...SIGLENT 136 SDS2000X E User Manual...
Page 200: ...SIGLENT 174 SDS2000X E User Manual Figure 95 Built in Arb interface...
Page 206: ...SIGLENT 180 SDS2000X E User Manual Figure 100 WiFi setting menu...
Page 215: ...SIGLENT SDS2000X E User Manual 189 Figure 108 Option Information...