History
The history function can record the waveforms of the input channels before pressing the
Run/Stop
button. In the run state, the oscilloscope records input waveforms continuously.
After the memory reaches the maximum frame, the new frames will cover the old frames.
The first frames acquired are the first to be covered.. in a circular buffer configuration.
To use the History function, the horizontal format must be set to
YT
.
Do the following steps to record and replay waveform:
1. Press the
History
button on the front panel to enable the History function.
When in the run state, the waveform will enter the stop state.
When in the stop state, and then enable the History function, the oscilloscope will
stay in the stop state.
Press
History
or
Stop
button to turn off the History function.
2. Press the
List
softkey to turn on or off the list display. The list records the timestamp
of every frame. It is accurate to microseconds.
Figure 70 History
3.
Press the
Frame
softkey; then turn the
Universal Knob
to select the frame to display.
The format of the
Frame
is A/B; A is the current frame number and B is the
maximum frame number you can set.
The maximum frame number is determined by the current sampling point
(Current
value) and sample rate.
When pressing the
Run/Stop
button or enabling the history function, you may not
Summary of Contents for SDS2000X-E Series
Page 1: ...User Manual SDS2000X E Series Digital Oscilloscope UM0102E E01A SIGLENT TECHNOLOGIES CO LTD...
Page 14: ......
Page 69: ...SIGLENT SDS2000X E User Manual 43 Figure 13 x Interpolation Figure 14 Sin x x Interpolation...
Page 98: ...SIGLENT 72 SDS2000X E User Manual Figure 29 Relative Window Trigger...
Page 100: ...SIGLENT 74 SDS2000X E User Manual Figure 30 Interval Trigger...
Page 103: ...SIGLENT SDS2000X E User Manual 77 Figure 32 State Dropout Trigger...
Page 105: ...SIGLENT SDS2000X E User Manual 79 Figure 33 Runt Trigger...
Page 125: ...SIGLENT SDS2000X E User Manual 99 Figure 49 CAN Trigger...
Page 129: ...SIGLENT SDS2000X E User Manual 103 Figure 51 LIN Trigger...
Page 162: ...SIGLENT 136 SDS2000X E User Manual...
Page 200: ...SIGLENT 174 SDS2000X E User Manual Figure 95 Built in Arb interface...
Page 206: ...SIGLENT 180 SDS2000X E User Manual Figure 100 WiFi setting menu...
Page 215: ...SIGLENT SDS2000X E User Manual 189 Figure 108 Option Information...