Note:
There are two indicators on the side of the SLA1016, one is red (Power), another
is green (Ready)
The red indicator: If SLA1016’s power is normal, this indicator turns on, or
else it blinks or turns off
.
The green indicator: When the SLA1016 is connect to the oscilloscope, this
indicator turns on for seconds then turns off, and then blinks, and finally keeps
the light on. When data is transferred between the SLA1016 and the
oscilloscope, the indicator blinks.
Connect the Digital Probes
1. If necessary, turn off the power supply to the device-under-test.
Turning off the power to the device-under-test prevents damage that might occur
if you accidentally short lines together while connecting probes. You can leave the
oscilloscope powered on because no voltage appears at the probes.
2. Connect one of the digital probe cable ports to the digital channels connector on the
front panel of the SDS2000X-E, and the other one to the digital channels connector of
the digital probe.
The digital probe cable is keyed so you can connect it only one way. You do not
need to power-off the oscilloscope to insert the probe.
3. Connect a fly line to one of the digital probe ground pins, connect a grabber to the fly
line, and then connect the grabber to a ground trace/pin from the device-under-test.
The ground lead improves signal fidelity to the oscilloscope, ensuring accurate
measurements.
4. Connect a fly line to one of the digital probe digital channel pins; connect a grabber to
the fly line, and then connect the grabber to a node in the circuit you want to test.
5. Repeat step 4 until you have connected all points of interest.
Summary of Contents for SDS2000X-E Series
Page 1: ...User Manual SDS2000X E Series Digital Oscilloscope UM0102E E01A SIGLENT TECHNOLOGIES CO LTD...
Page 14: ......
Page 69: ...SIGLENT SDS2000X E User Manual 43 Figure 13 x Interpolation Figure 14 Sin x x Interpolation...
Page 98: ...SIGLENT 72 SDS2000X E User Manual Figure 29 Relative Window Trigger...
Page 100: ...SIGLENT 74 SDS2000X E User Manual Figure 30 Interval Trigger...
Page 103: ...SIGLENT SDS2000X E User Manual 77 Figure 32 State Dropout Trigger...
Page 105: ...SIGLENT SDS2000X E User Manual 79 Figure 33 Runt Trigger...
Page 125: ...SIGLENT SDS2000X E User Manual 99 Figure 49 CAN Trigger...
Page 129: ...SIGLENT SDS2000X E User Manual 103 Figure 51 LIN Trigger...
Page 162: ...SIGLENT 136 SDS2000X E User Manual...
Page 200: ...SIGLENT 174 SDS2000X E User Manual Figure 95 Built in Arb interface...
Page 206: ...SIGLENT 180 SDS2000X E User Manual Figure 100 WiFi setting menu...
Page 215: ...SIGLENT SDS2000X E User Manual 189 Figure 108 Option Information...