SIG
L
ENT
SDS2000X-E User Manual
27
Unit
Select the amplitude display unit for the current channel. The available units are volts
denoted by “
V
” and amps shown as “
A
”. When the unit is changed, the unit displayed in
the channel label will change accordingly.
1. Press
CH1
button on the front panel to enter the CH1 function menu.
2. Press the
Next Page
softkey to enter the second page of the CH1 function menu.
3. Press the
Unit
softkey to select the desired unit
V
or
A
.
The default setup is
V.
Deskew
Deskew allows you to add or subtract a mathematical time delay to a channels time data.
This can be helpful in coordinating signals that may have different cable lengths or other
causes of time skew between channels.
The valid deskew range of each analog channel is ±100 ns.
1. Press
CH1
button on the front panel to enter the CH1 function menu.
2. Press the
Next Page
softkey to enter the second page of the CH1 function menu.
3. Press the
Deskew
softkey. Then turn the
Universal Knob
to change deskew.
Invert
When
Invert
is set to
On
, the voltage values of the displayed waveform are inverted.
Invert affects how a channel is displayed and it keeps the trigger settings.
Inverting a channel also changes the result of any math function selected and measure
function.
1. Press
CH1
button on the front panel to enter the CH1 function menu.
2. Press the
Next Page
softkey to enter the second page of the CH1 function menu.
3. Press the
Invert
softkey to turn on or off the invert display.
Summary of Contents for SDS2000X-E Series
Page 1: ...User Manual SDS2000X E Series Digital Oscilloscope UM0102E E01A SIGLENT TECHNOLOGIES CO LTD...
Page 14: ......
Page 69: ...SIGLENT SDS2000X E User Manual 43 Figure 13 x Interpolation Figure 14 Sin x x Interpolation...
Page 98: ...SIGLENT 72 SDS2000X E User Manual Figure 29 Relative Window Trigger...
Page 100: ...SIGLENT 74 SDS2000X E User Manual Figure 30 Interval Trigger...
Page 103: ...SIGLENT SDS2000X E User Manual 77 Figure 32 State Dropout Trigger...
Page 105: ...SIGLENT SDS2000X E User Manual 79 Figure 33 Runt Trigger...
Page 125: ...SIGLENT SDS2000X E User Manual 99 Figure 49 CAN Trigger...
Page 129: ...SIGLENT SDS2000X E User Manual 103 Figure 51 LIN Trigger...
Page 162: ...SIGLENT 136 SDS2000X E User Manual...
Page 200: ...SIGLENT 174 SDS2000X E User Manual Figure 95 Built in Arb interface...
Page 206: ...SIGLENT 180 SDS2000X E User Manual Figure 100 WiFi setting menu...
Page 215: ...SIGLENT SDS2000X E User Manual 189 Figure 108 Option Information...