38
SDS2000X HD Service Manual
129.51M, 148.24M, 169.68M, 194.23M, 222.33M, 254.49M,
291.30M, 333.44M, ,381.67M, 436.88M, 500.00M
SDS2204X HD
50 Ω
1M, 1.14M, 1.31M, 1.50M, 1.72M, 1.97M, 2.25M, 2.57M,
2.95M, 3.37M, 3.86M, 4.42M, 5.06M, 5.79M, 6.63M, 7.59M,
8.69M, 9.94M, 11.38M, 13.03M , 14.91M, 17.07M, 19.54M,
22.36M, 25.60M, 29.30M, 33.54M, 38.39M, 43.94M, 50.30M,
57.58M, 65.90M, 75.43M, 86.35M, 100.00M, 113.14M,
129.51M , 148.24M, 169.68M, 200.00M, 222.32M, 254.49M,
291.30M , 350.00M
SDS2104X HD
50 Ω
1M, 1.14M, 1.31M, 1.50M, 1.72M, 1.97M, 2.25M, 2.57M,
2.95M, 3.37M, 3.86M, 4.42M, 5.06M, 5.79M, 6.63M, 7.59M,
8.69M, 9.94M, 11.38M, 13.03M , 14.91M, 17.07M, 19.54M,
22.36M, 25.60M, 29.30M, 33.54M, 38.39M, 43.94M, 50.30M,
57.58M, 65.90M, 75.43M, 86.35M, 100.00M, 113.14M,
129.51M , 148.24M, 169.68M, 200.00M
6.
It is recommended to adjust the timebase to 1 us/div.
7.
Record the
“ Stdev ” measurement of the waveform as V
stdev
and the first frequency point as
reference value as V
ref.
8.
Set the vertical scale of the oscilloscope channel under test to the next setting according to Table
10, and repeat step 3 to step 7. The amplitude of the 9500B output is always 6*vertical scale.
9.
Check the other channels in the same way as step 1 to step 8.
10.
Disconnect the test connection.
After the test, calculate the dB value ( regarding the amplitude at 1 MHz as 0 dB ) at every frequency
point. Check if the value is in the limited range in Table 12 Use the formula below to calculate the dB
value:
dB value =
20 ∗ log
10
Vstdev
Vref