SDS2000 X HD Ser vice Manual
47
11.
Test at 2 V/div and 200 mV/div, using the same procedure as steps 9 to 10.
12.
Check the other analog channels in the same way as steps 8 to 11.
13.
Disconnect the test connection
When checking the EXT trigger channel, it is the same as checking analog channels. Record the
reading displays on the screen of the 9500B, and check to verify it is within the specified range in