48
SDS2000X HD Service Manual
6
Disassembly Procedures
This chapter describes how to remove major parts from the SDS2000X HD series oscilloscope.
6.1
Safety Consideration and Cautions
Only qualified personnel should perform the disassembly procedures. Disconnect the power before
you begin to remove or replace the parts. Otherwise, potential personal injuries or damages to the
components may occur.
WARNING
HAZARDOUS VOLTAGES
Maintenance is performed with the power supply on and without protective
covers. Only qualified personnel with proper tools and protection should
perform maintenance. Perform maintenance with the power cord
disconnected whenever possible.
WARNING
AVOID ELECTRICAL SHOCK
Hazardous voltages exist on the power supply and LCD modules. To avoid
electrical shock, disconnect the power cord and wait at least three minutes for
the capacitors to discharge before you begin disassembly.
CAUTION
ESD CAUTION
Electrostatic discharge ( ESD ) sensitive devices inside. Electronic
components may be damaged without detection. Proper ESD precautions
should be taken to avoid performance degeneration or loss of functionality. At
a minimum, place the oscilloscope on a properly grounded ESD mat and wear
a properly grounded ESD strap.
CAUTION
REMOVE POWER TO AVOID DAMAGE
Remove power before you start to disassemble or replace parts. DO NOT
disassemble or replace parts with the power on. Damage to the components
may occur.