SDS2000 X HD Ser vice Manual
55
6.3.7 To Remove the Probe interface board and Mainboard
Steps:
1.
Remove the one PM3*6 screws on the probe interface board, unplug the connecting cable and
remove the probe interface board.
2.
Remove the four nuts from the front BNC connectors.
3.
Remove the thirteen PM3*6 screws on the mainboard and remove the mainboard.