SDS2000 X HD Ser vice Manual
61
8.2
Required Equipment
Table 16 Required Equipment
Equipment
Critical Specifications
Recommended Model
Digital Multi-meter
Accuracy ± 0.1%
10 mV resolution
SIGLENT SDM3065X or
Agilent 34401A
Oscilloscope
200 MHz BW
1 MΩ impedance
SIGLENT SDS1204X-E
8.3
Troubleshooting Flowchart
The following flowchart describes how to troubleshoot the oscilloscope in the most general case. This
does not guarantee a 100% recovery of all possible hardware failures. Contact SIGLENT if you cannot
solve the problem.
Begin
Test Power
Supply Module
Good?
Replace Power
Supply Module
Test Main Board
Good?
Replace Main
Board
Test LCD
Good?
Replace LCD
NO
YES
NO
YES
NO
End
YES