Note
: In Auto mode, if the signal satisfies the trigger conditions but cannot
trigger the oscilloscope stably, it may be that interval between two trigger
events exceeds the timeout period. Try Normal mode in this case.
Normal
:
Triggers and acquisitions only occur when the trigger conditions are
met. Otherwise, the oscilloscope holds the last waveform on the display and
waits for the next trigger. Normal mode is suitable for acquiring:
Only events specified by the trigger settings
Rare events
Single
:
Captures and displays a single frame which satisfies the trigger
conditions, and then stops. The following trigger events are ignored until the
Single acquisition is restarted. Single mode is suitable for:
One-shot events or periodic signals, such as power-on\off waveforms
on a power rail
Rare events
15.5
Trigger Type
15.5.1 Overview
The trigger modes of the SDS5000X are digital designs. Compared with analog
trigger circuits, digital triggers can not only greatly optimize trigger precision,
trigger jitter, but also support multiple trigger types and complex trigger
conditions.
Summary of Contents for SDS2000X Plus
Page 1: ...SDS5000X Series Digital Oscilloscope User Manual UM0105X E01E...
Page 2: ......
Page 158: ...SDS5000X Series Digital Oscilloscope User Manual 156 WWW SIGLENT COM Ack Acknowledge bit...
Page 165: ...SDS5000X Series Digital Oscilloscope User Manual WWW SIGLENT COM 163...
Page 240: ...SDS5000X Series Digital Oscilloscope User Manual 238 WWW SIGLENT COM...
Page 248: ...SDS5000X Series Digital Oscilloscope User Manual 246 WWW SIGLENT COM...
Page 361: ...SDS5000X Series Digital Oscilloscope User Manual WWW SIGLENT COM 359...