SDS5000X Series Digital Oscilloscope User Manual
WWW.SIGLENT.COM
109
TV Standard
Scan Type
Sync Pulse
NTSC
Interlaced
Bi-level
PAL
Interlaced
Bi-level
HDTV 720P/50
Progressive
Tri-level
HDTV 720P/60
Progressive
Tri-level
HDTV 1080P/50
Progressive
Tri-level
HDTV 1080P/60
Progressive
Tri-level
HDTV 1080i/50
Interlaced
Tri-level
HDTV 1080i/60
Interlaced
Tri-level
Custom
The table below shows the parameters of the custom video trigger:
Frame Rate
25Hz, 30 Hz, 50 Hz,60
Hz
Of Lines
300 ~ 2000
Of Fields
1,2,4,8
Interlace
1:1, 2:1,4:1,8:1
Trigger Position
Line
Field
(line value)/1(1:1)
1
(line value)/2 (2:1)
1,2, 3, 4,5,6,7,8
(line value)/4(4:1)
1,2, 3, 4,5,6,7,8
(line value)/8(8:1)
1,2, 3, 4,5,6,7,8
Line value: The number of lines set in the
Of Lines
(300 ~ 2000).
In the custom video trigger type, the corresponding "Of Fields" varies with the
selection of
the “Interlace” ratio. Therefore, the number of fields selected and
the number of lines corresponding to each field can also be varied. If the "Of
Summary of Contents for SDS2000X Plus
Page 1: ...SDS5000X Series Digital Oscilloscope User Manual UM0105X E01E...
Page 2: ......
Page 158: ...SDS5000X Series Digital Oscilloscope User Manual 156 WWW SIGLENT COM Ack Acknowledge bit...
Page 165: ...SDS5000X Series Digital Oscilloscope User Manual WWW SIGLENT COM 163...
Page 240: ...SDS5000X Series Digital Oscilloscope User Manual 238 WWW SIGLENT COM...
Page 248: ...SDS5000X Series Digital Oscilloscope User Manual 246 WWW SIGLENT COM...
Page 361: ...SDS5000X Series Digital Oscilloscope User Manual WWW SIGLENT COM 359...