SDS5000X Series Digital Oscilloscope User Manual
114
WWW.SIGLENT.COM
If the upper trigger level is within the waveform amplitude range while
the lower trigger level is out of the waveform amplitude range, the
oscilloscope will trigger on the rising edge only.
If the lower trigger level is within the waveform amplitude range while
the upper trigger level is out of the waveform amplitude range, the
oscilloscope will trigger on the falling edge only.
To set window trigger via the Absolute window type
Refer to "Adjust Upper/Lower Level" in the section " Slope Trigger".
To set window trigger via the Relative window type
When the window trigger type is set to
"Relative", touch the trigger descriptor
box. The pop-up menu will show two
user defined
parameters: “Level +/-
Delta" and "Level Center".
The above two parameters can be set
in the following two ways:
1. Select the parameter in the
Level +/-Delta
area of the quick menu, then
set the parameter value by the virtual keypad, or the Level knob on the
front panel. To set the center level is the similar.
2. Directly use the
Level
knob on the front panel. Press the knob to switch
between
“Level +/-Delta" and "Center Level", and rotate it to set values.
Summary of Contents for SDS2000X Plus
Page 1: ...SDS5000X Series Digital Oscilloscope User Manual UM0105X E01E...
Page 2: ......
Page 158: ...SDS5000X Series Digital Oscilloscope User Manual 156 WWW SIGLENT COM Ack Acknowledge bit...
Page 165: ...SDS5000X Series Digital Oscilloscope User Manual WWW SIGLENT COM 163...
Page 240: ...SDS5000X Series Digital Oscilloscope User Manual 238 WWW SIGLENT COM...
Page 248: ...SDS5000X Series Digital Oscilloscope User Manual 246 WWW SIGLENT COM...
Page 361: ...SDS5000X Series Digital Oscilloscope User Manual WWW SIGLENT COM 359...