SDS5000X Series Digital Oscilloscope User Manual
122
WWW.SIGLENT.COM
15.5.14
Setup/Hold Trigger
Clock source and data source need to be set in setup/hold trigger setting. The
set-up time starts when the data signal crosses the trigger level and ends when
the specified clock edge arrives. The holding time starts when the specified
clock edge arrives and ends when the data signal crosses the trigger level again
(as shown in the figure below). When the set-up time or hold time meets the
preset time limit conditions, the oscilloscope will trigger.
Touch the
CLK Setting
region to set the clock source, threshold and slope;
Touch the
Data Setting
region to set the data source, threshold and state.