15.8
Trigger Coupling
The coupling setting of trigger is only valid when the trigger source is C1~C4,
EXT or EXT/5.
DC: All of
the signal’s frequency components are coupled to the trigger
circuit for high frequency bursts or where the use of AC coupling would shift
the effective trigger level.
AC: The signal is capacitively coupled. DC levels are rejected. See the
datasheet for details of the cut-off frequency.
HFR: Signals are DC coupled to the trigger circuit and a low-pass filter
network attenuates high frequencies (used for triggering on low
frequencies). See the datasheet for details of the cut-off frequency.
LFR: The signal is coupled through a capacitive high-pass filter network,
DC is rejected and low frequencies are attenuated. For stable triggering on
medium to high frequency signals. See the datasheet for details of the cut-
off frequency.
15.9
Noise Reject
Noise Reject adds additional hysteresis to the trigger circuitry. By increasing
the trigger hysteresis, the noise immunity becomes better but the trigger
sensitivity degrades.
Summary of Contents for SDS2000X Plus
Page 1: ...SDS5000X Series Digital Oscilloscope User Manual UM0105X E01E...
Page 2: ......
Page 158: ...SDS5000X Series Digital Oscilloscope User Manual 156 WWW SIGLENT COM Ack Acknowledge bit...
Page 165: ...SDS5000X Series Digital Oscilloscope User Manual WWW SIGLENT COM 163...
Page 240: ...SDS5000X Series Digital Oscilloscope User Manual 238 WWW SIGLENT COM...
Page 248: ...SDS5000X Series Digital Oscilloscope User Manual 246 WWW SIGLENT COM...
Page 361: ...SDS5000X Series Digital Oscilloscope User Manual WWW SIGLENT COM 359...