SDS5000X Series Digital Oscilloscope User Manual
WWW.SIGLENT.COM
165
A. Frame Type: Both, CAN, CAN FD
B. When the "Frame Type" is CAN FD, turn on or
off ISO
C. When the "Frame Type" is CAN FD, set the
BRS (Bit Rate Switch) to Both, Normal or FD
D. Trigger Condition: Start, Remote, ID, ID +
Data, Error
E. Return to previous menu
Trigger Condition
Start
— The oscilloscope triggers at the beginning of the frame.
Remote
— The oscilloscope triggers on a remote frame with a specified ID.
ID, ID Bits (11-bit or 29-bit) and Curr ID Byte (1st, 2nd, 3rd, or 4th byte) can
be set. Curr ID Byte is used to specify the byte to be adjusted when using
the universal knob.
ID
— The oscilloscope triggers on the data frame that matches the specified
ID. ID, ID Bits (11-bit or 29-bit), and Curr ID Byte (1st, 2nd, 3rd, or 4th byte)
can be set.
ID + Data
— The oscilloscope triggers on the data frame that matches the
specified ID and data.ID, ID Bits (11-bit or 29-bit), Curr ID Byte (1st, 2nd,
3rd, or 4th byte), Data1 and Data2 can be set.
Error
— The oscilloscope triggers on the error frame.
Error Frame
Stuff Bit Error
Summary of Contents for SDS2000X Plus
Page 1: ...SDS5000X Series Digital Oscilloscope User Manual UM0105X E01E...
Page 2: ......
Page 158: ...SDS5000X Series Digital Oscilloscope User Manual 156 WWW SIGLENT COM Ack Acknowledge bit...
Page 165: ...SDS5000X Series Digital Oscilloscope User Manual WWW SIGLENT COM 163...
Page 240: ...SDS5000X Series Digital Oscilloscope User Manual 238 WWW SIGLENT COM...
Page 248: ...SDS5000X Series Digital Oscilloscope User Manual 246 WWW SIGLENT COM...
Page 361: ...SDS5000X Series Digital Oscilloscope User Manual WWW SIGLENT COM 359...