SDS5000X Series Digital Oscilloscope User Manual
166
WWW.SIGLENT.COM
CRC Mismatch Error: The oscilloscope triggers when the
calculated CRC does not match the transmitted CRC.
Stuff Bit Cnt Err: Effective only when ISO is turned on, the
oscilloscope triggers when the stuff bit count is incorrect.
Stuff Bit Cnt Par. Err: Effective only when ISO is turned on, the
oscilloscope triggers when the polarity of the stuff bit count is
incorrect.
16.8.3 CAN FD Serial Decode
The configuration of CAN FD decoding is similar to that of I2C decoding.
On the bus:
ID is displayed in green.
BRS (Bit Rate Switch) is displayed in light yellow.
ESI (Error State Indicator) is displayed in blue.
L (Data Length) is displayed in light yellow.
D (Data) is displayed in white.
CRC is displayed in blue.
Ack is displayed in pink.
The red point at the end of a segment indicates there is not enough
space on the display to show complete content of a frame and some
content is hidden.
In the list view: