SDS5000X Series Digital Oscilloscope User Manual
178
WWW.SIGLENT.COM
If you set the trigger condition to Slow Channel:
ID
can be selected in the hexadecimal range of
0x0 to 0xF (short serial / enhanced serial with 4
bits ID
)
or 0x00 to 0xFF
(
enhanced serial with 8
bits ID
)
. If it is selected as "0xXX", the ID is
ignored.
Compare Type
can be specified as Equal, Not
Equal, Less than, Less or Equal, Greater than,
Greater or Equal, in Range of, out of Range.
Data
can be selected in the hexadecimal range
of 0x00 to 0xFF (short serial
)
or 0x0000 to
0xFFFF
(
enhanced serial with 4 bits ID
)
or 0x000
to 0Xfff(enhanced serial with 8 bits ID). If it is
selected as "0xXX", the data is ignored.
Error
— the oscilloscope triggers on the error frame. Errors include Successive
Sync Pulses Error
,
Pulse Period Error, Fast Channel CRC Error, Slow Channel
CRC Error, All CRC Errors.
Successive Sync Pulses Error: triggers on a sync pulse whose width varies
from the previous sync pulse's width by greater than 1/64 (1.5625%, as
defined in the SENT specification).
Pulse Period Error: triggers if a nibble is either too wide or too narrow (for
example, data nibble < 12 (11.5) or > 27 (27.5) ticks wide). Sync, S&C, data,
or checksum pulse periods are checked.
Fast Channel CRC Error
:
triggers on any Fast Channel Message CRC error.
Summary of Contents for SDS2000X Plus
Page 1: ...SDS5000X Series Digital Oscilloscope User Manual UM0105X E01E...
Page 2: ......
Page 158: ...SDS5000X Series Digital Oscilloscope User Manual 156 WWW SIGLENT COM Ack Acknowledge bit...
Page 165: ...SDS5000X Series Digital Oscilloscope User Manual WWW SIGLENT COM 163...
Page 240: ...SDS5000X Series Digital Oscilloscope User Manual 238 WWW SIGLENT COM...
Page 248: ...SDS5000X Series Digital Oscilloscope User Manual 246 WWW SIGLENT COM...
Page 361: ...SDS5000X Series Digital Oscilloscope User Manual WWW SIGLENT COM 359...