SDS5000X Series Digital Oscilloscope User Manual
WWW.SIGLENT.COM
209
18.6
Measurement Statistics
Touch
Statistics Setting
in the measure dialog box to recall the statistics config
dialog box:
A. Set the maximum number of
samples for the statistics function.
The setting range is 0 ~ 1024, or
unlimited. If there is no limit, the
number of statistics will be
accumulated. If there is a limit,
when the maximum number of
statistics N is reached, only the
latest N measurements will be
counted. When count is limited,
only the first period measurement
value of each frame is counted