SDS5000X Series Digital Oscilloscope User Manual
WWW.SIGLENT.COM
221
Setting
Offset
in the integral menu provides an approach to correct the DC
offset of the source. Small DC offsets in the input signal (or even small offset
errors of the oscilloscope itself) may cause the integral output waveform to
"ramp" up or down, as shown below:
Integral without Offset
Integral with Offset
In addition, the integral operation can be performed within a specified gate.
Touch
Gate
area in the math dialog box, and enable
Gate
function, then set
Gate A
and
Gate B
to define the gate. The setting of the gate cursors is similar
to that of normal cursors. See " Select and Move Cursors" for details.