SDS5000X Series Digital Oscilloscope User Manual
228
WWW.SIGLENT.COM
Select
Operation
as FFT in the math dialog box and
touch
Config
to recall the configuration dialog box:
A. Set the maximum points (2
n
, n = 11 ~ 21)
B. Set
the
window
type
(Rectangle,
Blackman,
Hanning,
Hamming
and
Flattop)
C. Automatically set vertical scale, reference
level, center frequency and horizontal
scale
according
to
the
maximum
frequency component in the previous
frame
D. Select the display mode (Split, Full Screen
and Exclusive)
E. Select the FFT mode (Normal, Average
and Max-Hold)
F. Set the average count in average mode
G. Reset average
H. Return to previous menu
Windows
Spectral leakage in FFTs can be considerably decreased when a window is
used. SDS5000X provides five windows which have different characteristics
and are applicable to different scenarios.
For example, for a two-tone signal with a very close frequency interval, it is