dynamic range limitation when trying to measure weak voltage and (or) current
in the presence of relatively large switching voltage and (or) current.
To correct the skew between the oscilloscope channels or probes, the deskew
procedure should be performed once initially, and re-run when any part of the
hardware setup changes (for example, a different probe, different oscilloscope
channel, etc.) or when the ambient temperature changes. The deskew
procedures by using the DF2001A deskew fixture is as followed.
A. Demagnetize and zero-adjust the current probe
B. Set up connections to the DF2001A deskew fixture
Small Loop
Large Loop
Available current
probes
CP4020 (100 kHz,20
Arms)
CP4050 (1 MHz,50 Arms)
CP5150 (12 MHz,150
Arms)
CP5500 (5 MHz,500 Arms)