SDS5000X Series Digital Oscilloscope User Manual
WWW.SIGLENT.COM
8 1
14.2
Acquisition Setup
14.2.1 Overview
Touch
Acquire Menu
on the quick menu of the timebase settings, or press the
Acquire button on the front panel, or touch the menu bar
Acquire
>
Menu
to recall
the Acquire dialog box on the right side.
A. Select the acquisition mode
(Normal/Peak/Average/ERES)
B. Select the maximum memory depth
C. Select the interpolation mode
D. Select the Acq mode
Acquisition:
See the section " Acquisition" for details.
Memory Depth
: The maximum memory depth that can be supported.
According to the formula "acquisition time = sample points x sample interval",
setting a larger memory depth can achieve a higher sample rate for a given
timebase, but more samples require more processing time, degrading the
waveform update rate. With 250 Mpts memory depth, SDS5000X can still run
at full sample rate (5 GSa/s) even when set to the 5 ms/div timebase.
Note
: The memory depth here is the upper limit of the memory space allocated
by the oscilloscope. The actual sample points is related to the current timebase
Summary of Contents for SDS2000X Plus
Page 1: ...SDS5000X Series Digital Oscilloscope User Manual UM0105X E01E...
Page 2: ......
Page 158: ...SDS5000X Series Digital Oscilloscope User Manual 156 WWW SIGLENT COM Ack Acknowledge bit...
Page 165: ...SDS5000X Series Digital Oscilloscope User Manual WWW SIGLENT COM 163...
Page 240: ...SDS5000X Series Digital Oscilloscope User Manual 238 WWW SIGLENT COM...
Page 248: ...SDS5000X Series Digital Oscilloscope User Manual 246 WWW SIGLENT COM...
Page 361: ...SDS5000X Series Digital Oscilloscope User Manual WWW SIGLENT COM 359...