34 SDS6000A Quick Start
Calibration
The oscilloscope is calibrated at the factory prior to being shipped. The calibration is
run at 23 °C (± 2 °C) and is valid for temperatures 23 ± 5 °C. Within this temperature
range, the oscilloscope will meet all specifications once warmed up.
Warm up the oscilloscope for at least 20 minutes prior to each use or calibration in
order for it to reach a stable operating temperature. Specifications are not guaranteed
during the warm-up period.
Whenever the oscilloscope is used in an environment outside 23 ± 5 °C, or when it
has been more than one month since the previous calibration, manual calibration is
recommended. To perform a self-calibration:
Touch Utility
>
Menu
>
Maintenance
>
Self Calibration
or
>
Maintenance
>
Self Calibration