111 SNA5000A Vector Network Analyzer User Manual
Limit table:
⚫
Show table: Check this box to open a table at the bottom of the screen to create and
edit limits.
⚫
Load table: Call the saved limit table, The saved format of the limit table is CSV by default.
Global pass/fail:
Pass/fail indicator is a convenient way to monitor all measurement status.
Global pass/fail on
- Check this box to display the global pass / fail status.
Policy
- Select which of the following requirements must be met for the global pass/fail
status to display "pass":
⚫
All tests must pass
- This setting will read the results from the limit test. If all tests
(limit test is on) pass, the global pass/fail status will be "pass".
⚫
All measurements must pass
- Unless all measured data points are within the
established test limits and the limit test is "on", the
global pass/fail status will be "fail".
Note:
In this mode, if there is a test that does not set the limit test to the on state, the global
pass/fail will display "fail".
Limit line setting
Click
Math→Analysis→Limit Table→Limit
to open the limit line setting interface, we
can also click
Math→Analysis→Limit→Show Table
to open the limit line setting interface,
as shown in the figure below.
Figure 4-7 Limit line setting interface
Limit line type:
⚫
Unused: Invalid limit line.
⚫
Max: Above this limit line, the measurement fails.
⚫
Min: Below this limit line, the measurement fails.
⚫
Delete: Delete limit line.
Begin stimulus:
Frequency of the limit line’s starting point.
End stimulus:
Frequency of the limit line’s ending point.
Begin response:
Response of the limit line’s starting point.