141 SNA5000A Vector Network Analyzer User Manual
Min
:
Selects 3 acquisitions.
Better
:
Selects 5 acquisitions.
Max
:
Selects 6 acquisitions.
Note
: Only 3 sets of data can be acquisitions at most when the frequency point is below 40MHz.
6.3.2
Processing setup
Press the “AvgBw” on the front panel,prss AvgBW > SA to enter the menu interface of
processing setup menu.
Figure 6-5 the interface of processing setup menu
6.3.2.1
RBW
Set the RBW (resolution bandwidth) in order to distinguish between signals which have
frequency components that are near one another.
●
Reducing the RBW will increase the frequency resolution. but lead to a longer sweep time
(influenced by both RBW and VBW when the sweep time is in automatic mode).
●
RBW varies with the Span/RBW setting in Auto RBW mode.