SNA5000A Vector Network Analyzer User Manual 46
windows share the same scale settings.
•
When window or full coupling is enabled, the scale settings for the working trace will be
adopted by other coupling traces of the same format
•
If traces of a different format exist, all traces of that format will be set as the least numbered
trace in that format
•
Once enabled, the scale settings for all coupling traces of the same format can be changed
with any coupling trace in the working state
Selected Windows:
Available when selecting a window or all methods. The selected window participates in
scaling coupling. By default, all windows are selected. Cancel the check box to disable zoom
coupling for this window.
About "Auto Scaling" and "Scaling Coupling":
Automatic scaling using the coupling method affects the work trace in the work window.
All traces coupled to this trace will be set with the new scale of the working trace. This will
cause some traces to not appear on the screen.
Full automatic scaling using the coupled method
•
Close: All traces in the working window are automatically scaled separately.
•
Window: All traces in each selected window are automatically scaled to conform to a
common set of scaling factors.
•
All: All traces in all selected Windows are automatically scaled to conform to a common
set of scaling factors.
2.7.3
Electrical delay
Electrical delay is a mathematical function that simulates the variable length of a lossless
transmission line. Linear phase shifts through the device can be compensated using the
electrical delay function. Use this feature to determine only the linear phase deviation of the
device. Electrical delay can be set separately for each measuring trace.
Operating steps:
Press
Scale
, use the knob or arrow key to bring focus to the
Electrical Delay
parameter
item. Press
ENTER
or Knob to enter the editing state. After the editing is completed, move the
cursor to the specified position by up and down direction keys or Knob. Press
ENTER
or Knob
to select the current option.
Delay time:
Specifies the value of the delay added or removed, expressed in time or distance.
This compensates for the linear phase shift through the device. Electrical delay
can be set separately for each measuring trace.