3 SNA5000A Vector Network Analyzer User Manual
Protection level: Class I
Note:
Installation (overvoltage) category Class II
indicates the local supply level is
suitable for equipment connected to the AC power supply.
Installation (overvoltage) category Class I indicates the signal levels suitable
for terminals connected to the RF source.
Pollution level Class II indicates it only occurs in a dry and non-conductive
environment, sometimes we should consider the temporary conductivity caused
by concentration.
Protection level Class I indicates grounding equipment, it prevents electric
shock by connecting the equipment to the ground wire.
CAUTION:
Do not apply excessive pressure or strike the surface of the touch screen.
CAUTION:
Do not exceed the maximum voltage marked on the front panel connectors.
Cooling requirement
This instrument is cooled through a built-in fan. To keep adequate ventilation,
a gap of at least 15 cm should be left on both sides, as well as the front and rear
panels of the instrument.
CAUTION:
Do not block the vents located along the side and real panel of the instrument.
CAUTION:
Do not let any external objects enter the instrument through the vents.
AC power supply
The instrument accepts 100-240V, 50/60/400Hz AC power. The maximum
power consumption is 90W with complete options.
Note:
The instrument can operate within the following input ranges:
Voltage range:
90 - 264 Vrms
90 - 132 Vrms
Frequency range:
47 - 63 Hz
380 - 420 Hz
Power supply and grounding
The instrument has a three-terminal plug and an IEC320 (Type C13)
standard connector for the input power and grounding connections. The
grounding terminal on the socket is directly connected to the instrument shell. To
prevent electric shock, the plug must be inserted into a well-grounded socket.
Please use the power cord provided to connect the instrument to the power
source.