55 SNA5000A Vector Network Analyzer User Manual
Required calibration kits: Short circuit, open circuit, load, defined through or flush through
Corrected system error:
•
Directivity (Source port)
•
Isolation
•
Frequency response
transmission tracking (receiver
port)
•
Source match (source port)
•
Load matching (receiver port) is
used only to generate
transmission trace items
•
Frequency response reflection
tracking (source port)
One port (reflection):
Application: Used to accurately calibrate any test port suitable for reflection measurement only
Overall accuracy: High
Required calibration kits: Short circuit, open circuit, and load or ECAL module
Corrected system error:
•
Directivity
•
Frequency response reflection
tracking
•
Source match
Open circuit response/short circuit response:
Application: Quick calibration of any test port suitable for reflection measurement only
Overall accuracy: Low
Calibration kits required: Open circuit or short circuit.
Corrected system error:
•
Frequency response reflection tracking
Direct/Transport Tracking (isolation optional):
Application: Quick calibration of any pair of test ports suitable for transmission measurements
only. Isolation calibration is generally not recommended.
Overall accuracy: Low
Required calibration kit: Through
Isolation: Each load corresponds to a vector network analyzer test port.
Corrected system error:
Summary of Contents for SNA5000A Series