59 SNA5000A Vector Network Analyzer User Manual
3.3.1.3
Standards Tab
Allows you to Add, Edit, or Delete cal standards in a cal kit.
1) In the "Standard" tab, click "Add" to add the Standard sample.
Figure 3-4 add standard interface
2) Standards dialog box help, the following fields apply to ALL standard types.
Figure 3-5 The help dialog of the Standard
A. Identification
Standard ID: Number in list of standards
Label: Type of standard. This usually appears in prompts for standards.
Description: Description of standard.
B. Connector
Indicates the type and gender (Male, Female, None) of the standard.
Thru and Isolation standards have two connectors.
C.Frequency Range
Min: Defines the lowest frequency at which the standard is used for calibration.
Max: Defines the highest frequency at which the standard is used for calibration.