67 SNA5000A Vector Network Analyzer User Manual
Figure 3-16 View the newly created Cal Kit
3.3.2
S Parameter calibration wizard
S parameter user calibration can be started after selecting the required S parameter
calibration type in the calibration wizard and setting the corresponding parameters.
Operating instructions:
Press the
Cal
key, use the knob or the arrow key to make the focus fall on
Basic Cal
, and
press
Enter
to enter the setting interface of the calibration kit:
Figure 3-17 Calibration wizard interface
Select the number of ports to be calibrated from the port menu in the calibration wizard
interface, and select the corresponding calibration kit model in the Cal Kit pull-down option.
Select the gender properties of the calibration connector from the DUT Connector drop-down