71 SNA5000A Vector Network Analyzer User Manual
1.
Preset
Vector network analyzer.
2. Set internal source output power, IF bandwidth, scan points, test frequency band, and other
parameters.
3. Select
Cal
→
Basic Cal
to enter the setting interface.
4. Select only
Port1
or
Port2
, select the desired calibration kit in the
Cal Kit
, select the Type
of calibration connector,
Female
or
Male
, in the
DUT Connector
, select
OSL (Open, Short, Load)
in the
Cal Type
, and click
Next
to proceed to the next step.
•
Connect the OPEN calibration element to the test port according to the interface
prompts:
•
Click
Open
for calibration
•
Then connect the SHORT calibration element to the test port, click
Short
for
Calibration.
•
Finally, connect the LOAD calibration element to the test port, click the
Load
for
Calibration.
•
After the calibration, click
Finish
to exit the calibration interface, complete the
calibration, and save the calibration data.
3.3.6
Transmission response calibration (two ports)
Two-port transmission response calibration is to connect the THRU calibration kit between
two ports to calculate the transmission tracking error in the test device error model. If the LOAD
calibration kit is used for isolation calibration at the same time, isolation error (crosstalk error)
can also be calculated.
Figure 3-23 Schematic diagram
Operating steps:
1.
Preset
Vector network analyzer.
2. Set internal source output power, IF bandwidth, scan points, test frequency band, and other
parameters.
3. Select
Cal
→
Basic Cal
to enter the setting interface.
4. Select both
Port1
and
Port2
, select the desired calibration kit in the
Cal Kit
, select the Type