73 SNA5000A Vector Network Analyzer User Manual
Figure 3-24 Error model diagram
Operating steps:
1.
Preset
vector network analyzer.
2. Set internal source output power, IF bandwidth, scan points, test frequency band, and other
parameters.
3. Select
Cal
→
Basic Cal
to enter the Settings screen.
4. Select both
Port 1
and
Port 2
, select the desired calibration kit in the
Cal Kit
, select the
type of calibration connector,
Female
or
Male
, in the
DUT Connector
, select
SOLT
in the
Cal Type
, and click
Next
to proceed to the next step.
4.
According to the interface prompts, connect the Open, Short, and Load calibration
elements to Port 1 successively, and click the corresponding
Open
,
Short,
and
Load
items successively to complete the Port 1 calibration.
6. According to the interface prompts, connect the Open, Short, and Load calibration elements
to Port2 successively, and click the corresponding
Open
,
Short,
and
Load
items
successively to complete the Port 2 calibration.
7. According to the interface prompts, connect the THRU calibration kit between the two ports
to perform the
Thru
calibration.
8. After the calibration, click
Finish
to exit the calibration interface, complete the calibration,
and save the calibration data.
3.3.9
SOLR unknown through calibration (two ports)
In contrast to the two-port SOLT calibration, the error term in the test unit can be calculated
without knowing the S-parameter of the THRU straight-through calibration. The calibration
method is similar to that of SOLT in that it is possible to connect the LOAD to both ports for
isolated calibration.