77 SNA5000A Vector Network Analyzer User Manual
3.5 Receiver calibration
The vector network analyzer receiver power has been factory calibrated, but the factory
calibration may not guarantee sufficient accuracy in every situation. The user can conduct a
receiver power calibration before the test to improve the test accuracy. To ensure the
calibration accuracy of the receiver, it is better to calibrate the power of the internal source
before the calibration of the receiver.
Figure 3-27 Receiver calibration setting interface
Operating steps:
1.
Preset
vector network analyzer.
2. Set internal source output power, IF bandwidth, scan points, test frequency band, and other
parameters.
3. Select
Cal
→
Receiver Cal
to enter the Settings screen.
4. Select the test channel, select the internal source output port, select the receiver port
5. Connect the two ports using an appropriate specification RF cable.
6.
Set parameters such as power offset.
7.
Calibrate R
or
Calibrate T
or
Calibrate Both
can be clicked to begin receiver calibration.
8.
Click
OK
to complete the calibration and exit the receiver calibration interface.