87 SNA5000A Vector Network Analyzer User Manual
Figure 3-36 Adapter removal and insertion functions
Adapter removal operation steps:
1. Connect the test cable of the vector network analyzer to the adapter, and configure the
parameters of the network analyzer. First, perform the full 2-port calibration.
2. Select
Cal
→
Adapter Removal
to go to the Settings screen and select
Remove
in the
Adapter
tab.
3. Remove the adaptor from the test cable, Select
Cal
→
Adapter Removal
→
Calibration
to
enter the
Adapter Removal
configuration screen. Select the port connected to the adaptor
to be removed, select the calibration kit type, and select the polarity of the calibration kit.
Click
Next
to connect the Open, Short and Load calibration elements in turn for full 1 port
calibration.
4. Click
Finish
to complete the calibration and click Adapter Removal to ON to complete the
Adapter Removal operation.
Summary of Contents for SNA5000A Series