91 SNA5000A Vector Network Analyzer User Manual
Figure 3-40 Confidence Check
2.
Compare the data-trace with the memory trace and verify that the measurement is correct.
3. The steps to compare when both data and memory traces are displayed are as below:
a) Press “Math” on the front panel,Press “Math” > Memory > Math > Data + Memory.
b) Press “Scale” on the front panel,Press”Scale” > Scale > Auto Scale.
c) Determine if the differences between trace lines are acceptable.
3.9.4
Orientation
The ECal module automatically detects the connection between SNA's test ports and ECal
module's ports. You can turn off this function to set ports manually.
Manual port setting procedure for Full 2-Port calibration
a) Press “Cal” on the front panel, Press “Cal” > ECal > Orientation > Manual.
b) Specify a test port of the VNA.
c) Specify an ECal port for the port of the VNA you specified.