95 SNA5000A Vector Network Analyzer User Manual
Figure 4-1 Data format interface
5. Marker coupling:
The newly created marker turns off coupling by default. At this time,
the marker is only set on a single trace, and the markers on different
traces do not affect each other.
If we need multiple markers on different
traces to share settings, we can use the marker coupling function. Click
the
Coupled
command, and select the range of the marker coupling in
the pop-up menu for coupling operation.
When the marker coupling
mode is on, the markers of the same number and X-axis (such as
frequency or time) on all traces in the coupling range share the settings,
and the marker’s setting on the current trace will be used by default.
If
the current marker is not in the coupling range, the marker with the same
number on the trace with the smallest trace number is used.
Note: when the marker coupling mode is on, the marker on the trace
within the coupling range will automatically move to the same X-axis
coordinate.
6. Marker table:
The newly created marker closes the marker table by default. If we need to
display all marker data in the form of a table, we can use the marker table