SNA5000A Vector Network Analyzer User Manual 90
Figure 3-39 ECal config menu
A
.
Select the number of ports to calibrate and configure the port number. Using ECal, you
can perform 1-Port Reflection calibration, Full 2-Port calibration, Full 3-Port calibration,
and Full 4-Port calibration.
B
.
Select the Cal Type.
C
.
When "Orientation" is Auto, click "Next" to enter the automatic alignment directly. When
"Orientation" is Manual, click "Next" to specify the calibration port and go to auto-
calibration.
3.9.3
Confidence Check
1.
Press “Cal“ on the front panel, press Cal > ECal > Confidence Check to enter the menu
interface of Confidence Check.
Summary of Contents for SNA5000A Series