SSA3000X Plus User Manual 99
4.
The spectrum lines on the screen do not update for a long period of time:
(1) Check whether the screen is locked; if so, press
Esc
to unlock it.
(2) Verify whether all the trigger conditions have been met and whether there is a valid trigger signal
inputting.
(3) Check whether the analyzer is in single sweep.
(4) Check whether the current sweep time is too long.
5.
Wrong measurement results or poor precision:
To calculate the system errors and check the measurement results and precision, refer to the
introductions in “
Specifications
”. To reach these specifications, please:
(1) Check whether all the external devices are successfully connected and are working normally.
(2) Get some knowledge of the signal under measurement and set appropriate instrument parameters.
(3) Make measurements under proper conditions; for example, warm-up the instrument appropriately
and operate the instrument under the specified environment temperature.
(4) Calibrate the instrument regularly to reduce or avoid errors that might occur over time.
If you need a specific calibration after the stated calibration period, contact
SIGLENT
or get paid
service from authorized measurement agencies.
6.
System Messages:
The instrument may display prompt messages, error messages or state messages according to the
current working status. These messages are displayed to help you to use the instrument correctly
rather than instrument failures.
#
User system message
Message on screen
System message description(1~199)
1
SWT_OOR
Sweep time out of range
2
RBW_OOR
RBW out of range
19
PEAK_UNFOUND
No peak found, please change the search config
20
IMD_FREQ_OOR
Frequency of intermodulation products is out of range
21
AUTO_FAIL
No peak found
22
EXT_REF_PLUG_IN
EXT ref plug in
23
EXT_REF_PLUG_OUT
EXT ref plug out
150
LAN_PLUG_IN
Ethernet cable plug in
151
LAN_PLUG_OUT
Ethernet cable plug out
152
IP_CONFLICT
IP address conflict
153
IP_INVALID
IP address invalid
154
NETM_INVALID
Netmask address invalid
155
GWAY_INVALID
Gateway address invalid
Device error(600~799)
605
LIC_INVALID
License invalid
606
ADC_ERROR
Warning, ADC overload