1.7 Firmware Operation
1.7.1 Check System Information
Users can get the system information by press
System
-
>“
System Info
”, including
Product Model, Serial and Host ID
Software Version and hardware Version
Option Information
1.7.2 Load Option
Refer to the procedures below to activate the options you have purchased.
1.
Press
System
-
>“
System Info
”->“
Load Option
”
2.
Enter the license key in the onscreen window. Press
Enter
to confirm your input and terminate the
license key input. Or
3.
Load the .lic file provided by pressing
File
-
>“
Load
” from internal memory or USB stick.
The option will be enabled after rebooting.
1.7.3 Firmware
Upgrade
Follow this procedure to update the instrument firmware:
1.
Download the firmware package from an official SIGLENT website,
http://www.siglent.com/ens
,
2.
Extract and copy the .ADS file into the root directory of an USB stick.
3.
Plug the USB stick into the USB Host connector. Press
System
->
“
System Info
”-> “
Firmware
Update
”; find the .ADS file in USB stick.
4.
Press the
“
Load
”, the analyzer will perform the update process automatically.
The upgrade process will take several minutes. When the upgrade is completed, the machine will
reboot.
Any interruption during the update process will result in update failure and system data loss. This
is
not covered under the warranty
and the user will bear repair costs and shipping.
Do not remove the USB storage device until the update is finished.
1.8 Touch Operation
The analyzer has a 10.1 inch multi-touch screen and supports various gesture operations. Including:
Press or click on the upper-right-corner of the screen to enter the main menu;
Swipe up and down or left and right in the waveform area to change the X-axis center coordinate
or Y-axis reference level;
Perform two-points scaling in the waveform area to change the X-axis span;