SSA3000X Plus User Manual 47
2.3.2 Marker ->
1.
M->CF
Set the center frequency of the analyzer to the frequency of the current marker.
If the
Normal
marker is selected, the center frequency will be set to the frequency of the current
marker.
If the
Delta
or
Delta Pair
marker is selected, the center frequency will be set to the frequency of
the Delta Marker.
The function is invalid in Zero span.
2.
M -> CF Step
Set the center frequency step of the analyzer to the frequency of the current marker.
If the
Normal
marker is selected, the center frequency step will be set to the frequency of the
current marker.
If the
Delta
or
Delta
Pair marker is selected, the center frequency step will be set to the frequency
of the Delta Marker.
The function is invalid in Zero span.
3.
M -> Start Freq
Set the start frequency of the analyzer to the frequency of the current marker.
If the
Normal
marker is selected, the start frequency will be set to the frequency of the current
marker.
If the
Delta
or
Delta Pair
marker is selected, the start frequency will be set to the frequency of the
Delta Marker.
The function is invalid in Zero span.
4.
M -> Stop Freq
Set the stop frequency of the analyzer to the frequency of the current marker.
If the
Normal
marker is selected, the stop frequency will be set to the frequency of the current
marker.
If the
Delta
or
Delta Pair
marker is selected, the stop frequency will be set to the frequency of the
Delta Marker.
The function is invalid in Zero span.
5.
M ->Ref Level
Set the reference level of the analyzer to the amplitude of the current marker.
If the
Normal
marker is selected, the reference level will be set to the amplitude of the current
marker.
If the
Delta
or
Delta Pair
marker is selected, the reference level will be set to the amplitude of the
Delta Marker.
6.
ΔM->Span
Set the span of the analyzer to the frequency difference between the two markers in Delta
marker type.
If
the Normal
marker is selected, this function is invalid.