SSA3000X Plus User Manual 67
Gauss
Half Sine
Off
Note: The rule of common filter type selected
Transmitter Filter
Measure Filter
Reference Filter
Sqrt Nyquist
Sqrt Nyquist
Nyquist
Nyquist
Off
Nyquist
Gauss
Off
Gauss
Half Sine
Off
Half Sine
3. Filter Alpha/BT
For Sqrt Nyquist and Nyquist filter settings, you can set the alpha parameter. It can be set the
same as the transmitter.
For a Gauss filter, use the BT parameter. It can be set the same as the transmitter.
Table 3-5 Filter Parameter
Parameter
Explanation
Default
0.35
Range
0 ~ 1
Unit
1
Knob Step
0.01
Direction Key Step
0.1
4. Filter Length
Set the symbols number of the filter selected. It can be set the same as the transmitter.
Table 3-6 Filter Length
Parameter
Explanation
Default
64
Range
2 ~ 128
Unit
1
Knob Step
1
Direction Key Step
5
3.2.1.6
Statistic
1. Statistic
Turn on the statistical function, the measurement result will show the maximum and minimum of
statistics. Turn off the statistical function, the measurement result will only show the real-time
measurement value. Statistical function is turned off by default.