70 SSA3000X Plus User Manual
measure the lower modulation frequency signal. The EqLPF bandwidth should be as narrow as
possible to improve the S/N ratio, but also need to be larger than the modulation frequency.
Table 3-9 EqLPF
Parameter
Explanation
Default
IFBW/6
Range
Off, IFBW/6, IFBW/20, IFBW/60, IFBW/200, IFBW/600, IFBW/2000
Unit
kHz
、
Hz
3.2.2.3
Average
Open and close the average option for the measurement result. It can set the average number. When
the Avg Number is set to off, the column title "Average" in numerical results view will be changed to
"Current". The "Average" measurement result will be stable if the average number is set to larger
values.
Table 3-10 Avg Number
Parameter
Parameter
Default
10
Range
1 ~ 1000
Unit
1
Knob Step
1
Direction Key Step
10
3.2.2.4
Restart Meas
After this function is performed, the statistics will be cleared and restarted. If the average function is
turned on, the average calculation of measurement results will be cleared and restarted as well.
3.3 Sweep and Functions
3.3.1 Trigger
Press
Trigger
to open the menu. The analyzer will begin a sweep only with the selected trigger
conditions are met. A trigger event is defined as the point at which your trigger source signal meets the
specified trigger level.
Free Run: New sweep starts as soon as possible after the current sweep ends.
RF Trigger: Activates the trigger condition that starts the next sweep if the detected RF envelops
voltage rises to a level set by the RF trigger level.
External: The trigger even is the rising or falling edge of the external trigger signal.