SIGLENT
34 SSG5000X Programming Guide
3.6.1.3
Frequency Offset ([:SOURce]:FREQuency:OFFSet)
Command
Format
[:SOURce]:FREQuency:OFFSet <freq>
[:SOURce]:FREQuency:OFFSet?
Instruction
Set the frequency offset of a downstream circuit element
Get the frequency offset of a downstream circuit element
Parameter
Type
Float, unit: Hz, kHz, MHz, GHz, Default “Hz”
Parameter
Range
-200 GHz ~ 200 GHz
Return
Float, unit: Hz
Default
0 Hz
Menu
FREQ > Freq Offset
Example
FREQuency:OFFSet 2 MHz
3.6.1.4
Phase Offset ([:SOURce]:PHASe)
Command
Format
[:SOURce]:PHASe <phase>
[:SOURce]:PHASe?
Instruction
Set the phase of the RF output signal
Get the phase of the RF output signal
Parameter
Type
Float, unit: deg
Parameter
Range
-360 deg ~ 360 deg
Return
Float, unit: deg
Default
0 deg
Menu
FREQ > Phase Offset
Example
PHASe 20
3.6.1.5
Phase Reset ([:SOURce]:PHASe:RESet
[:SOURce]:PHASe:REF)
Command
Format
[:SOURce]:PHASe:RESet
[:SOURce]:PHASe:REF