SIGLENT
SSG5000X Programming Guide 37
AUTO
Internal level control is activated/deactivated automatically depending on
the operating state.
Return
Enumeration
Default
AUTO
Menu
LEVEL > ALC State
Example
POWer:ALC ON
3.6.2.5
Flatness List State ([:SOURce]:CORRection[:FLATness])
Command
Format
[:SOURce]:CORRection[:FLATness] ON|OFF|1|0
[:SOURce]:CORRection[:FLATness]?
Instruction
Activate/deactivate flatness correction list.
Parameter
Type
Boolean
Parameter
Range
ON|OFF|1|0
Return
Boolean
Default
0
Menu
LEVEL > Flatness
Example
CORRection:FLATness ON
3.6.2.6
Flatness List Add Row
([:SOURce]:CORRection:FLATness:PAIR)
Command
Format
[:SOURce]:CORRection:FLATness:PAIR <freq>,<power>
Instruction
Insert a new row in the flatness list.
Parameter
Type
Float, Float
Parameter
Range
Freq: Full freq range
Power: Full power range
Return
None
Default
None
Menu
LEVEL > Flatness > [+]